Application and Technical Data
Linearity
Linearity is defined as the maximum deviation of the calibration
curve (an average of the upscale and downscale readings) from
a straight line so positioned as to minimize the maximum deviation.
Platinum resistance elements have a nearly linear output while
nickel and nickel-iron (Balco) sensors are quite curved. Copper
elements are also nearly linear over their narrow temperature
range.
Stability
Stability is the relationship of a sensor's original resistance curve
to its curve after being in service. Drift rates published by a manufacturer
must be assumed to be applicable to high purity laboratory
environment probes. The published drift rates of 0.0°C are to
be considered general and not necessarily quantitative.
Several parameters affect stability in a platinum sensor used in
industrial processes. Thermal and mechanical treatment cause
physical changes in the crystalline structure of the platinum causing
different resistances at different temperatures. Chemical reactions
involving platinum and impurities as well as migration of
internal materials can affect a sensor output. A shunting effect
due to insulation resistance deterioration is another influencing
occurrence.
The drift caused by these conditions is not normally catastrophic
except in rare instances. Attempts to establish a statement of stability
in industrial applications would result in an ambiguous
approximation at best.
Self-Heating
Since an RTD measures temperature by passing a current
through a resistor (the RTD), the error known as self-heating
occurs.
Primarily the sensor's mass, its internal construction, the measurement
current and to a large degree environmental conditions
determine the magnitude of this error. Normally a very small current,
usually 1-5 milliamps is used in the excitation circuit to minimize
this joule heating of the sensor. Thermo Sensors' internal
construction technique maximizes heat transfer quality to further
reduce the effect.
An installation condition requiring large mass hardware such as
thermowells or protective tubes coupled with an environment of
still or slow moving air is going to experience a great deal more
self-heating than the next example. a small diameter (.250" O.D.)
direct immersion probe mounted in an environment of flowing
water (min. 3 ft./sec) could totally dissipate the error.
Fortunately if a small measuring current (1-2 ma) is used, selfheating
errors will be well within acceptable levels for industrial
applications.
To approximate the amount of error; consider that normally the
dissipation constant will be of the magnitude of 20-100 mw/0°C,
and use the following formula.
| Self-heating error = |
Power |
| Dissipation constant |
| Example: |
Measurement current - 2 ma
Resistance of sensor - 140 ohms
dissipation constant - 50 mw/0°C
|
Power = 12R
= (.002)2 (140) = 0.56 MW
| Error = |
.56 mw |
.011°C |
| 50 mw/0°C |
Time Constant
Time constants are values used to indicate the time it takes a sensor
to read 63.2% of a step change in temperature. This test is
conducted in water flowing at 3 ft/sec or 20 ft/sec in air. Typically
this measurement is made by plunging a sensor at room temperature
into a bath at 80°C and noting the time required to reach
63.2% of that step change. Generally speaking, it takes approximately
five (5) time constants before 100% of the step change is
realized.
Several variables affect the response time of sensors. Diameter of
the sheath, material of the sheath and internal construction for different
temperature ranges are the most variable. It is possible,
however, to approximate the time constant for a particular group
of sensors based on diameter and assuming the sheath material
is a 300 series stainless steel.
| These approximations are: |
| .125" |
1.1sec. |
| .188" |
1.7sec. |
| .250" |
2.2sec. |
Note: elements capable of a lower range of -250°C (to +600°C)
have similar time constants.
These time constants should serve only as a general approximation
for direct immersion sensors. Sensors installed in thermowells,
protection tubes or that are mounted in conditions allowing
appreciable stem losses are not subject to even these general
constants.
In the rare instance where the response time absolute needs to
be known; response time testing must be conducted to provide a
time constant.
Insulation Resistance
To prevent an unacceptable shunting effect between the sensing
element and the probe sheath, care must be taken to assure good
insulation quality.
In all sensors and particularly those in industrial service, high
temperature operation, contamination and moisture absorption
are potential problems.
To eliminate the effects of these occurrences, Thermo Sensors
adheres to stringent manufacturing procedures. Reliatemp's
Insulation Resistance will always be > 2000 megaHMS at or
below 100°C.
Repeatability
By definition repeatability of a sensor is the relationship of the
original resistance at 0°C and any different resistance at 0°C after
being subjected to the following test.
The sensor shall be brought slowly to the upper limits of its temperature
range and then exposed to air at room temperature. It
shall then be brought slowly to its lower limit, and exposed to air
at room temperature.
This procedure is repeated ten times. The resistance of 0°C is
then measured and the difference from the pre-testing resistance
is 0°C is noted.
For a typical platinum probe, the resistance should not change
more than 0.3°C for a 0.12% sensor or 0.15°C for a 0.06% sensor.
The 0.12% and 0.06% are original resistance tolerances at
0°C of the element.
|